SPEEDFLOW® ULTRA

Speedflow ULTRA is particularly suitable for filtration in every application in which an “ultrapure” fluid has to be obtained under higher flow rate and for large volumes.

Speedflow ULTRA filter could work as a barrier to endo- toxins and bacterial contaminants prior to IV medical administration and for filtration of:

  • Large volumes aqueous IV solutions
  • Drug preparation in pharmaceutical applications
  • TPN preparation
  • Ultrafiltration in on-line HDF

FEATURES

> Two vents for faster air elimination during priming and re-priming operation

> Free of latex, PVC, and substances of animal origin.


SPECIFICATIONS

Biosafety

Materials comply with USP class VI-121°C test, ISO 10993, REACH, RoHS


Dimensions

WxLxD: 59.0 x (75.0-100.0) x 25.4 mm

Weight:35.8 gr


Filtration Area

60.00cm2 PES

22.00 cm2 PTFE


Inlet/Outlet Connectors Inlet: tube ID 4.8 x OD 6.8 Outlet: tube ID 4.8 x OD 6.8

Other tubing connections available upon request.


Materials Of Construction

Filter media: Hydrophilic HI-FLO PES 0.2 µm and 0.2 µm po- sitive charged, or 1.2 µm, and 5 µm

Vent: Hydrophobic PTFE membrane 0.03 µm Housing: Clear Modified Acrylic


Max Operating Pressure
2.0bar (29.00psi)

Operating Temperature
5-40°C

Minimum water flow rate
0.2 µm: 280ml/min at 80cm (31.5”) head pressure
1.2 µm: 1,08lt/min at 80cm (31.5”) head pressure
5.0 µm: 2,04lt/min at 80 cm (31.5”) head pressure
0.2 µm positive membrane: (pressure drop measured between inlet and outlet, flow on peristaltic pump)
200 ml/min at 80cm (31.5”) head pressure 420 ml/min at 255cm (100,4”) head pressure

Air flow
>400 cc/min@100mbar

Minimum water bubble point
0.2 µm: 3.7-4,8 bar (54-70psi)
1.2 µm: 0.7-1.0 bar (10-15psi)
5.0 µm: 0.15-0,3 bar (2.2-4.5psi)

Water breakthrough
PTFE Hydrophobic membrane: 3.0bar (43.5psi)

Pyrogenicity
<0.25 EU/ml using the LAL test method

Sterilization compatibility
EtO or Gamma (25kGy)

Bacterial retention
0.2 µm, 0.2 µm positive charged –>7LRV – Brevundimonas diminuta
1.2 µm - Candida albicans